EasyManua.ls Logo

R&S RTO6 - Page 144

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acquisition and waveform setup
R&S
®
RTO6
144User Manual 1801.6687.02 ─ 05
One filter is applied to a pair of channels. If you filter only on the input channels, you
can apply dif
ferent filters to the channels - one filter for channels 1 and 2 and another
filter for channels 3 and 4.
If you filter the trigger channel, the same filter must be used for the input channels to
ensure that all signals suit for analysis. The instrument of
fers only permitted combina-
tions and triggers on the filtered signal.
In high definition mode, digital filter settings are enabled automatically. You can change
the high definition bandwidth in the "Digital Filter" setup, which is applied to the chan-
nels.
Example:
"RF reject" for the trigger signal ensures that triggering is not caused by unexpected
glitches.
Channel BW limit (On/Off)
Enables the DSP filter for the correspondig input channel.
Remote command:
CHANnel<m>:
DIGFilter:
STATe on page 1383
Channel BW limit
If "Trigger BW limit" is "Off", it sets the limit frequency of the lowpass filter for input
channels. One filter is applied to a pair of channels.
If "Trigger BW limit" is "RF reject", it sets the limit frequency that is applied to the trig-
ger channel and to the input channels enabled for filtering.
Remote command:
CHANnel<m>:DIGFilter:CUToff on page 1383
TRIGger<m>:RFReject on page 1384
Vertical setup