Remote control commands
R&S
®
RTO6
1457User Manual 1801.6687.02 ─ 05
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
CALCulate:MATH<m>:ENVSelection <EnvelopeCurve>
Selects the upper or lower part of the waveform for mathematic calculation, or a combi-
nation of both.
The setting is relevant for waveforms with waveform arithmetic mode "Envelope"
or
with "Peak detect" decimation. All mathematic operations - except for derivation - can
be applied to envelope waveforms and waveforms with "Peak detect" decimation.
Suffix:
<m>
.
1..8
Selects the math waveform.
Parameters:
<EnvelopeCurve> MIN | MAX | BOTH
*RST: BOTH
Usage: Asynchronous command
Firmware/software: FW 2.25
CALCulate:MATH<m>:ARIThmetics <Arithmetics>
Selects the method to build the resulting math waveform from consecutive acquisitions.
The processing is similar to the waveform arithmetics - instead of the acquired wave-
forms the results of the mathematic formula are used to create envelope, average and
RMS.
To define the reset condition for arithmetics calculation, use ACQuire:
ARESet:MODE.
Suffix:
<m>
.
1..8
Selects the math waveform.
Parameters:
<Arithmetics> OFF | ENVelope | AVERage | RMS | MINHold | MAXHold
W
aveform arithmetic mode
OFF
The math waveform is built according to the mathematic for-
mula.
ENV
elope
Detects the minimum and maximum math values in a sample
interval over a number of acquisitions.
Waveform analysis