EasyManua.ls Logo

R&S RTO6 - Page 1458

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1458User Manual 1801.6687.02 ─ 05
AVERage
Calculates the average from the math data of the current acqui-
sition and a number of acquisitions before. T
o define the number
of acquisitions, use ACQuire:COUNt.
RMS
The resulting math waveform is the root mean square of the cur-
rent acquisition and a number of acquisitions before. The result
is the average power spectrum. Number of acquisitions:
ACQuire:
COUNt
MAXHold
Determines the maximum result for each input value from the
math data of the current acquisition and a number of acquisi-
tions before. To define the number of acquisitions, use
ACQuire:COUNt.
MINHold
Determines the minimum result for each input value from the
math data of the current acquisition and a number of acquisi-
tions before. To define the number of acquisitions, use
ACQuire:COUNt.
*RST: OFF
Usage: Asynchronous command
CALCulate:MATH<m>:VERTical:OFFSet <V
erticalOf
fset>
Sets a voltage offset to adjust the vertical position of the math function on the screen.
Suffix:
<m>
.
1..8
Math waveform
Parameters:
<VerticalOffset> Negative values move the waveform au, positive values move it
down.
Range: -100E+12 to 100E+12
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
CALCulate:MATH<m>:VERTical:RANGe <VerticalRange>
Defines the range of FFT values to be displayed.
Suffix:
<m>
.
1..8
Math waveform
Waveform analysis

Table of Contents

Other manuals for R&S RTO6

Related product manuals