Parameters:
<StpAcqIdx> Index of the stop acquisition. The newest acquisition always has
the index "0".
Range: 0 to -(n-1)
Increment: 1
Usage: Asynchronous command
CHANnel<m>[:WAVeform<n>]:HISTory:TPACq <TimePerAcq>
Sets the display time for one acquisition. The shorter the time, the faster is the replay.
Suffix:
<m>
.
1..4
Selects the input channel.
<n> 1..3
Selects the waveform. If [:WA
Veform<n>] is omitted, waveform 1
is adressed.
Parameters:
<TimePerAcq> Range: 40E-6 to 10
Increment: 1
*RST: 0.05
Default unit: s
Usage: Asynchronous command
CHANnel<m>[:WAVeform<n>]:HISTory:PLAY
Starts and stops the replay of the history waveforms.
Suffix:
<m>
.
1..4
Selects the input channel.
<n> 1..3
Selects the waveform. If [:W
A
Veform<n>] is omitted, waveform 1
is adressed.
Usage: Event
Asynchronous command
CHANnel<m>[:WAVeform<n>]:HISTory:REPLay <AutoRepeat>
If ON, the replay of the history waveform sequence repeats automatically. Otherwise,
the replay stops at the stop index set with CHANnel<m>[:WAVeform<n>]:HISTory:
STOP.