Remote control commands
R&S
®
RTO6
1497User Manual 1801.6687.02 ─ 05
Usage:  Asynchronous command
MEASurement<m>:AMPTime:CLCK<n>:LSELect <ClkLevSel>
Selects the reference level of the clock on which the time is measured. Reference level
and clock slope define the time point for setup and hold measurements.
Suffix: 
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
<n> 1..2
The suffix is irrelevant, omit it.
Parameters:
<ClkLevSel> UPPer | MIDDle | LOWer
*RST:  MIDDle
Usage:  Asynchronous command
MEASurement<m>:AMPTime:DATA<n>:LSELect <DatLevSel>
Selects the reference level of the data on which the setup and hold time are measured.
Suffix: 
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
<n> 1..2
The suffix is irrelevant, omit it.
Parameters:
<DatLevSel>
UPPer | MIDDle | LOWer
*RST:  MIDDle
Usage:  Asynchronous command
MEASurement<m>:AMPTime:DTOTrigger<n>:SLOPe <DlyTrigSlp>
Sets the edge direction to be used for delay measurement.
Suffix: 
<m>
.
1..10
See 
"Selection of the measurement group: MEASurement<m>"
on page 1479.
<n> 1..2
The suffix is irrelevant.
Automatic measurements