Remote control commands
R&S
®
RTO6
1513User Manual 1801.6687.02 ─ 05
Parameters:
<ChPowBw> Range: 0 to 4E+9
Increment: 1
*RST: 0
Default unit: Hz
Usage: Asynchronous command
MEASurement<m>:SPECtrum:OBANdwidth <OccupiedBW>
Defines the percentage of the total power used to determine the occupied bandwidth.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Parameters:
<OccupiedBW> Range: 0.1 to 99.9
Increment: 1
*RST: 20
Default unit: %
Usage: Asynchronous command
MEASurement<m>:SPECtrum:CPOWer:CFRequency <ChPowCtrFreq>
Defines the center frequency from which the channel power is calculated over the
specified bandwidth.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Parameters:
<ChPowCtrFreq>
Range: 0 to 6E+9
Increment: 1
*RST: 0
Default unit: Hz
Usage: Asynchronous command
MEASurement<m>:SPECtrum:NDBDown <NDbDown>
Defines the threshold until which the samples to the left and right of the peak value are
analyzed in order to determine the "N dB down bandwidth".
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Automatic measurements