Remote control commands
R&S
®
RTO6
1567User Manual 1801.6687.02 ─ 05
Parameters:
<StartFreq> start frequency
Range: 0 to 5E+9
Increment: 1
*RST: 2E+9
Default unit: Hz
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:STOP <StopFreq>
Defines the stop frequency of the displayed frequency span.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<StopFreq> stop frequency
Range: 0 to 5E+9
Increment: 1
*RST: 2E+9
Default unit: Hz
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:CFRequency <CenterFreq>
Defines the position of the displayed frequency range, which is (Center - Span/2) to
(Center + Span/2). The width of the range is defined using the CALCulate:MATH<m>:
FFT:SPAN command.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<CenterFreq> center frequency
Range: 0 to 2E+12
Increment: 1
*RST: 2.5E+9
Default unit: Hz
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:FULLspan
Performs FFT calculation for the full frequency span.
Suffix:
<m>
.
1..8
math waveform