CALCulate:MATH<m>:FFT:PHASe:SCALe <PhaseScale>
Defines the scaling unit for phase display
.
Suffix: 
<m>
.
1..8
math waveform
Parameters:
<PhaseScale>
DEGRees | RADians
*RST:  DEGRees
Usage:  Asynchronous command
CALCulate:MATH<m>:FFT:PHASe:SUPPression <Suppression>
Enables noise suppression. Phase calculation is restricted to frequencies with a mini-
mum magnitude, the threshold value (see CALCulate:MATH<m>:
FFT:PHASe:
THReshold on page 1577).
Suffix: 
<m>
.
1..8
math waveform
Parameters:
<Suppression> ON | OFF
*RST:  OFF
Usage:  Asynchronous command
CALCulate:MATH<m>:FFT:PHASe:THReshold <SupprThres>
Defines the minimum frequency magnitude for which phases are calculated. This set-
ting is only available if CALCulate:
MATH<m>:
FFT:PHASe:SUPPression is set to
"ON".
Suffix: 
<m>
.
1..8
math waveform
Parameters:
<SupprThres> Range:  -180  to  180
Increment:  0.1
*RST:  0
Default unit: dBm
Usage:  Asynchronous command
CALCulate:MATH<m>:FFT:PHASe:UNWRap <Unwrap>
If enabled, phase shifts due to a limitation of the value range are eliminated.