Remote control commands
R&S
®
RTO6
1602User Manual 1801.6687.02 ─ 05
MTESt:EYEMask:TOFFset <MaskT
estName>,<RectDistT
op>
MTESt:EYEMask:TOFFset? <MaskTestName>
MTESt:EYEMask:BOFFset <MaskTestName>,<RectDistBtom>
MTESt:EYEMask:BOFFset? <MaskTestName>
Voltage distance from the eye shape center that limit the upper (TOFFset) and lower
(BOFFset) regions.
Parameters:
<RectDistBtom> Range: 0 to 100
Increment: 0.01
*RST: 0.5
Default unit: V
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:EYEMask:TBSYmmetric <MaskTestName>,<T
opBtomSymetrie>
MTESt:EYEMask:TBSYmmetric? <MaskTestName>
Sets bottom and top offsets to the same value so that the outer regions are symmetric
to the eye shape.
Parameters:
<TopBtomSymetrie> ON | OFF
*RST: ON
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:EYEMask:TBWidth <MaskTestName>,<RectangleLength>
MTESt:EYEMask:TBWidth? <MaskTestName>
Sets the time width of the outer regions, symmetric to the eye shape center.
Parameters:
<RectangleLength> Range: 0 to 100
Increment: 0.01
*RST: 0.5
Default unit: s
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
Mask testing