Remote control commands
R&S
®
RTO6
1601User Manual 1801.6687.02 ─ 05
MTESt:EYEMask:MSKRight <MaskT
estName>,<Copy>
MTESt:EYEMask:MSKRight?
<MaskTestName>
MTESt:EYEMask:MSKLeft <MaskTestName>,<Copy>
MTESt:EYEMask:MSKLeft? <MaskTestName>
Copies the eye shape to the right and left, respectively.
The distance of the copy is defined using MTESt:EYEMask:HPERiod.
Parameters:
<Copy> ON | OFF
*RST: OFF
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:EYEMask:HPERiod <MaskTestName>,<InterpattLength>
MTESt:EYEMask:HPERiod?
<MaskTestName>
Defines the time distance between the shape centers if MTESt:EYEMask:MSKLeft
and/or MTESt:EYEMask:MSKRight are ON.
Parameters:
<InterpattLength> Range: 0 to 100
Increment: 0.01
*RST: 0.5
Default unit: s
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:EYEMask:MSKTop <MaskTestName>,<Add>
MTESt:EYEMask:MSKT
op? <MaskTestName>
MTESt:EYEMask:MSKBottom <MaskTestName>,<Add>
MTESt:EYEMask:MSKBottom? <MaskTestName>
Enable the upper (top) and lower (bottom) mask region, respectively.
Parameters:
<Add> ON | OFF
*RST: OFF
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command