Remote control commands
R&S
®
RTO6
1690User Manual 1801.6687.02 ─ 05
Return values:
<FrameStart> Range: -100E+24 to 100E+24
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:I2C:FRAMe<n>:STOP?
Returns the end time of the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameStop> Range: -100E+24 to 100E+24
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:I2C:FRAMe<n>:SYMBol?
Returns the symbolic label of the specified frame if the label list is enabled.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the number of the frame in the current acquisition, 1...n.
Return values:
<Translation> String with symbolic name of the address
Usage: Query only
BUS<m>:I2C:FRAMe<n>:BCOunt?
Returns the number of bytes in the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Count> Byte count
Protocols