Remote control commands
R&S
®
RTO6
1744User Manual 1801.6687.02 ─ 05
DATA
T
riggers on a data word or data range on a specified channel or
on any channel.
T
o set the channel, use TRIGger<m>:I2S:TCONdition<n>:
CHANnel.
To set the data condition, use:
TRIGger<m>:I2S:TCONdition<n>:CONDtion,
TRIGger<m>:I2S:TCONdition<n>:DMIN and TRIGger<m>:
I2S:TCONdition<n>:DMAX.
WINDow
Triggers if the decoded data values stay inside a "window" that
is formed by a data range and a time specified by a number of
subsequent words. It considers a selected channel or all chan-
nels.
To set up a window trigger, you define the channel and data con-
dition in the same way as for DATA trigger type. Additionally, you
set the time limit with TRIGger<m>:I2S:SOWords.
CONDition
The frame condition trigger sets the trigger on an AND combina-
tion of data conditions on different channels. The instrument trig-
gers if all conditions are met inside one frame.
To set up a CONDition trigger, you define up to four channel and
data conditions in the same way as for DATA trigger type.
WSELect
WordSELect: Triggers on the selected edge of the WS line (I²S
standard, left- and right-justified). For TDM signals, it triggers on
the selected edge of the FSYNC line. Set the edge with
TRIGger<m>:I2S:WSSLope.
ECONdition
ErrorCONdition: Triggers on irregularities between the WS or
FSYNC edges.
*RST: DATA
Usage: Asynchronous command
TRIGger<m>:I2S:TCONdition<n>:CHANnel <Channel>
Selects the audio channel on which the instrument looks for the specified data condi-
tion.
Protocols