Remote control commands
R&S
®
RTO6
1759User Manual 1801.6687.02 ─ 05
BUS<m>:CMSB:NRZ:ENAPolarity <PolarityEnable>
Sets whether the transmitted enable signal is active when the voltage is below the
threshold (ENALow) or higher than it (ENAHIgh).
Suffix:
<m>
.
1..4
Parameters:
<PolarityEnable> ENALow | ENAHigh
*RST: ENAHigh
BUS<m>:CMSB:NRZ:POLarity <PolarityData>
Selects the polarity of the custom serial bus data signal in NRZ coding standards.
Suffix:
<m>
.
1..4
Parameters:
<PolarityData> ACTLow | ACTHigh
ACTLow
Active low: the value "1" is represented by a voltage below the
threshold
ACTHigh
Active high: the value "1" is represented by a voltage above the
threshold
*RST: ACTHigh
BUS<m>:CMSB:NRZ:THReshold:CLCK <ThresholdClock>
Sets the threshold for the clock signal digitization in the NRZ Clocked coding standard.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdClock> Range: -25 to 25
Increment: 0.1
*RST: 2
Default unit: V
BUS<m>:CMSB:NRZ:HYSTeresis:CLCK <HystClock>
Sets the hysteresis size on the clock channel.
Suffix:
<m>
.
1..4
Protocols