Remote control commands
R&S
®
RTO6
1760User Manual 1801.6687.02 ─ 05
Parameters:
<HystClock> Range: -25 to 25
Increment: 0.1
*RST: 0.1
Default unit: V
BUS<m>:CMSB:NRZ:THReshold:DATA <ThresholdData>
Sets the threshold for the data signal digitization in NRZ coding standards.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdData> Range: -25 to 25
Increment: 0.1
*RST: 2
Default unit: V
BUS<m>:CMSB:NRZ:HYSTeresis:DATA <HystData>
Sets the hysteresis size on the data line channel.
Suffix:
<m>
.
1..4
Parameters:
<HystData> Range: -25 to 25
Increment: 0.1
*RST: 0.1
Default unit: V
BUS<m>:CMSB:NRZ:THReshold:ENBLe <ThresholdEnable>
Sets the threshold for the enable signal digitization in NRZ coding standards.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdEnable> Range: -25 to 25
Increment: 0.1
*RST: 2
Default unit: V
BUS<m>:CMSB:NRZ:HYSTeresis:ENBLe <HystEnable>
Sets the hysteresis size on the enable line channel.
Suffix:
<m>
.
1..4