Remote control commands
R&S
®
RTO6
1761User Manual 1801.6687.02 ─ 05
Parameters:
<HystEnable> Range:  -25  to  25
Increment:  0.1
*RST:  0.1
Default unit: V
BUS<m>:CMSB:NRZ:THReshold:PRESet <PresetValue>
Sets the NRZ thresholds to predefined or individually definable voltage levels.
Suffix: 
<m>
.
1..4
Parameters:
<PresetV
alue> V05 | V2 | V5 | V7 | MANual
V05
Sets the clock and data threshold to +0.5 V (in case of NRZ
Unclocked: data threshold, only)
V2
Sets the clock and data threshold to +2.0 V (in case of NRZ
Unclocked: data threshold, only)
V5
Sets the clock and data threshold to +5.0 V (in case of NRZ
Unclocked: data threshold, only)
V7
Sets the clock and data threshold to +7.0 V (in case of NRZ
Unclocked: data threshold, only)
MANual
Allows to set individual threshold voltage levels
*RST:  V5
BUS<m>:CMSB:NRZ:THReshold:COUPling <ThresholdCoupling>
Couples the clock and data threshold values for the NRZ Clocked coding standard.
The values are coupled to the same number
.
Suffix: 
<m>
.
1..4
Parameters:
<ThresholdCoupling>
ON | OFF
ON
Activates coupling of the NRZ clock and data threshold values.
OFF
Disables coupling of the NRZ clock and data threshold values.
*RST:  ON