EasyManua.ls Logo

R&S RTO6 - Page 1772

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1772User Manual 1801.6687.02 ─ 05
Usage: Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMIN <DataMin>
TRIGger<m>:CMSB:ADV
anced:
DMIN <Frame>, <Field>, <Data>
Specifies the data pattern, or sets the start value of a data pattern range.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Field>
<Data>
Usage: Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DOPerator <DataOperator>
TRIGger<m>:CMSB:ADVanced:
DOPerator <Frame>, <Field>, <Operator>
Sets the operator for the data pattern.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Field>
<Operator>
OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
ANY = OFF
The data of the required pattern is not relevant.
EQUal | NEQual | L
THan | LETHan | GTHan | GETHan
This condition requires one endpoint value to be set using
TRIGger<m>:
CMSB:ADVanced:DMIN.
INRange | RANGe
This condition requires a range of endpoint values to be set
using TRIGger<m>:CMSB:ADVanced:DMAX and
TRIGger<m>:CMSB:ADVanced:DMIN.
Usage: Setting only
TRIGger<m>:CMSB:ADVanced:ERRor<n>:ENABle <Enable>
TRIGger<m>:CMSB:ADVanced:
ERENable <ErrorName>, <Enabler>
Enables triggering on errors. You can search for all error types in parallel.
Protocols

Table of Contents

Other manuals for R&S RTO6

Related product manuals