Remote control commands
R&S
®
RTO6
1807User Manual 1801.6687.02 ─ 05
TRIGger<m>:CAN:FDATa:DPTO <DataPositionT
o>
Sets the number of the last byte at which the required data pattern may start.
Parameters:
<DataPositionT
o> Range: 8 to 64
Increment: 1
*RST: 8
TRIGger<m>:CAN:ACKerror <AckError>
Triggers when the transmitter does not receive an acknowledgment - a dominant bit
during the Ack Slot.
The trigger type has to be set before:
TRIGger<m>:CAN:TYPE to ERRC.
Parameters:
<AckError> ON | OFF
*RST: ON
TRIGger<m>:CAN:BITSterror <BitStuffError>
T
riggers if astuff error occurs - when the 6th consecutive equal bit level in the men-
tioned fields is detected.
The trigger type has to be set before: TRIGger<m>:CAN:TYPE to ERRC.
Parameters:
<BitStuffError> ON | OFF
*RST: ON
TRIGger<m>:CAN:CRCerror <ChecksumError>
Triggers on CRC errors. A CRC error occurs when the CRC calculated by the receiver
dif
fers from the received value in the CRC sequence.
The trigger type has to be set before: TRIGger<m>:CAN:TYPE to ERRC.
Parameters:
<ChecksumError> ON | OFF
*RST: ON
TRIGger<m>:CAN:FORMerror <FormError>
Triggers when a fixed-form bit field contains one or more illegal bits.
The trigger type has to be set before:
TRIGger<m>:CAN:TYPE to ERRC.
Parameters:
<FormError> ON | OFF
*RST: ON
Protocols