Remote control commands
R&S
®
RTO6
1885User Manual 1801.6687.02 ─ 05
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:SIDData <SearchName>,<IdentifierData>
SEARch:TRIGger:FLXRay:SIDData
? <SearchName>
Enables the search for indentifier and data.
Parameters:
<IdentifierData>
ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay[:SSOFrame] <SearchName>,<FrameStart>
SEARch:TRIGger:FLXRay[:SSOFrame]? <SearchName>
Enables the search for a start of frame.
Parameters:
<FrameStart>
ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:SSYMbol <SearchName>,<CheckSymbol>
SEARch:TRIGger:FLXRay:SSYMbol? <SearchName>
Enables the search for specified symbol.
Parameters:
<CheckSymbol> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:CENable <SearchName>,<CycleCount>
SEARch:TRIGger:FLXRay:CENable? <SearchName>
Sets the operator to define a cycle count or a cycle count range.
Parameters:
<CycleCount>
OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
The cycle count is not relevant for the search condition.