Remote control commands
R&S
®
RTO6
1886User Manual 1801.6687.02 ─ 05
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one cycle count
number to be set with SEARch:
TRIGger:
FLXRay:CMIN.
INRange | OORange
In range, Out of range: Set the minimum and maximum value of
the range with SEARch:TRIGger:FLXRay:CMIN and SEARch:
TRIGger:FLXRay:CMAX.
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:CMAX <SearchName>,<CycleCountMax>
SEARch:TRIGger:FLXRay:CMAX? <SearchName>
Sets the the end value of a cycle count range if the condition SEARch:TRIGger:
FLXRay:
CENable is set to INRange or OORange.
Parameters:
<CycleCountMax> Range: 0 to 63
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:CMIN <SearchName>,<CycleCountMin>
SEARch:TRIGger:FLXRay:CMIN? <SearchName>
Specifies a cycle count - the number of the current FlexRay cycle - or sets the the start
value of an cycle count range.
Parameters:
<CycleCountMin> Range: 0 to 63
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:CSTep <SearchName>,<CycleCountStep>
SEARch:TRIGger:FLXRay:CST
ep?
<SearchName>
Specifies a step to search for each n-th cycle inside the given range. This allows for a
specific search if slot multiplexing is used.
The condition SEARch:TRIGger:FLXRay:CENable must be set to INRange or
OORange.