Loading...
Remote control commands
R&S
®
RTO6
1910User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the serial data bus.
Return values:
<Count> Total number of decoded frames.
Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:SENT:FRAMe<n>:STATus?
Returns the overall state of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial data bus.
<n> *
Selects the frame.
Return values:
<FrameState> OK | SYNC | PULSe | CRC | IRFL | FORM | INSufficient
OK: the frame is valid.
SYNC: Synchronization error occurred.
PULse: Pulse error occurred.
CRC: Cyclic redundancy check failed.
IRFL: Irregular frame length error occurred.
FORM: Format error occurred.
INSuf
ficient: The frame is not completely contained in the acqui-
sition. The acquired part of the frame is valid.
*RST
: OK
Usage: Query only
BUS<m>:SENT:FRAMe<n>:STARt?
BUS<m>:SENT:FRAMe<n>:ST
OP?
Returns the start time and stop time of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Protocols