Remote control commands
R&S
®
RTO6
1911User Manual 1801.6687.02 ─ 05
Return values:
<Start>, <Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:SENT:FRAMe<n>:CSValue?
Returns the CRC sequence value of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<ChecksumValue> To set the value format, use FORMat:
BPATtern
.
The values below range, increment and reset are decimal
values.
Range: 0 to 63
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:SENT:FRAMe<n>:DATA?
Returns the data of the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Parameters:
<Data> Comma-separated sequence of integer values (N, D1, D2,…,
DN). N is the number of nibbles in the frame, and D1…DN are
the values of the nibbles.
Example:
BUS:SENT:FRAMe4:DATA?
<-- 4,
3,15,11,9
Usage: Query only
BUS<m>:SENT:FRAMe<n>:IDTYpe?
Returns the identifier type of the selected frame.
Protocols