Remote control commands
R&S
®
RTO6
2065User Manual 1801.6687.02 ─ 05
TRIGger<m>:RFFE:DATA:DCON <DatPattOptor>
Sets the operator to trigger on a specific pattern or a range.
Parameters:
<DatPattOptor> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one pattern to
be set with
TRIGger<m>:RFFE:DATA:DMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:RFFE:DATA:DMIN and
TRIGger<m>:RFFE:DATA:DMAX
*RST: EQUal
Example: See "Triggering on RFFE bus"
on page 1274.
TRIGger<m>:RFFE:DATA:DMIN <DataPatternMin>
Specifies a pattern, or sets the start value of a pattern range.
Parameters:
<DataPatternMin> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (do not care).
Example: See
"Triggering on RFFE bus"
on page 1274.
TRIGger<m>:RFFE:DATA:DMAX <DataPatternMax>
Sets the end value of a range, if the data condition is set to INRange or
OORange. See
TRIGger<m>:RFFE:DATA:DCON.
Parameters:
<DataPatternMax> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (do not care).
Example: See
"Triggering on RFFE bus"
on page 1274.
TRIGger<m>:RFFE:DATA:ICONdition <DataIdxOperator>
Sets the operator to define the data frames in which the data pattern is expected.
Parameters:
<DataIdxOperator> EQUal | LTHan | LETHan | GTHan | GETHan | INRange |
RANGe
Protocols