Remote control commands
R&S
®
RTO6
2068User Manual 1801.6687.02 ─ 05
Parameters:
<MaskMin>
TRIGger<m>:RFFE:INTerrupt <Interrupt>
Sets the interrupt pattern for the interupt summary and notification trigger type.
Parameters:
<Interrupt>
TRIGger<m>:RFFE:ERRor:BP <BusParkError>
Enables the trigger on a bus park error - when an incorrect bus parlk cycle has been
found.
Parameters:
<BusParkError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:LENGth <LengthError>
Enables the trigger on a length error - when an incorrect length of the command
sequence has been found.
Parameters:
<LengthError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:NOResponse <NoResponse>
Enables the trigger on any No Response Frame.
Parameters:
<NoResponse> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:PARity <ParityError>
Enables the trigger on incorrect parity bits.
Parameters:
<ParityError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:SSC <SSCError>
Enables the trigger on an SSC error - when no valid SSC sequence has been found
after the idle time.