Remote control commands
R&S
®
RTO6
2081User Manual 1801.6687.02 ─ 05
Parameters:
<Interrupt> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page
1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
String parameter
Firmware/software: FW 3.30
SEARch:TRIGger:RFFE:ERRor:LENGth <SearchName>,<LengthError>
SEARch:TRIGger:RFFE:ERRor:LENGth? <SearchName>
Enables the search for length errors - when an incorrect length of the command
sequence has been found.
Parameters:
<LengthError> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName> String parameter
SEARch:TRIGger:RFFE:ERRor:PARity <SearchName>,<ParityError>
SEARch:TRIGger:RFFE:ERRor:PARity?
<SearchName>
Enables the search for incorrect parity bits.
Parameters:
<ParityError> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName> String parameter
SEARch:TRIGger:RFFE:ERRor:SSC <SearchName>,<SSCError>
SEARch:TRIGger:RFFE:ERRor:SSC? <SearchName>
Enables the search for SSC errors - when no valid SSC sequence has been found
after the idle time.
Parameters:
<SSCError> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName> String parameter