EasyManua.ls Logo

R&S RTO6 - Page 2108

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2108User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:HBTO:CRC:MAX <SearchName>,<CRCPatternMax>
SEARch:TRIGger:HBT
O:CRC:
MAX? <SearchName>
Sets the end value of the frame check range if SEARch:TRIGger:HBTO:CRC:
CONDition is set to INRange or OORange.
Parameters:
<CRCPatternMax> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:DATA:DCONdition <SearchName>,<DataOperator>
SEARch:TRIGger:HBT
O:DA
TA:DCONdition? <SearchName>
Sets the condition for the data. You can define an exact value or a value range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding SEARch:
TRIGger:HBTO:
DATA:DMIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with SEARch:TRIGger:HBTO:DATA:DMIN and
SEARch:TRIGger:HBTO:DATA:DMAX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:DATA:DMIN <SearchName>,<DataMin>
SEARch:TRIGger:HBTO:DA
TA:DMIN? <SearchName>
Specifies the data bit pattern, or sets the start value of a pattern range.
Parameters:
<DataMin> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
Protocols

Table of Contents

Other manuals for R&S RTO6

Related product manuals