EasyManua.ls Logo

R&S RTO6 - Page 211

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Triggers
R&S
®
RTO6
211User Manual 1801.6687.02 ─ 05
Pattern
The pattern contains the channel selection, and the logical operations structure of
hardware-based Boolean logic.
"Channel"
Select the channels to be considered. For qualification, you can
select all channel signals except for the trigger source. In pattern trig-
ger setup, the trigger source channel is selected by default, and you
can select all other channel signals.
"Boolean oper-
ator"
Defines the logical operation on the digital signal resulting from the
comparison with the trigger level.
"PASS"
: leaves the input value unchanged
"NOT": inverts the input value
Trigger types