Remote control commands
R&S
®
RTO6
2131User Manual 1801.6687.02 ─ 05
Parameters:
<ErrorCRC> ON | OFF
*RST: ON
TRIGger<m>:TBTO:ERRor:FEC <ErrorFEC>
Enables triggering, if more than 22 RS-FEC errors are detected.
Suffix:
<m>
.
1..3
Parameters:
<ErrorFEC> ON | OFF
*RST: ON
TRIGger<m>:TBTO:ERRor:OOR <ErrorOOR>
Enables triggering if values are out of range (OOR). For example, a preamble error, or
invalid symbols in control blocks.
Suffix:
<m>
.
1..3
Parameters:
<ErrorOOR>
ON | OFF
*RST: ON
TRIGger<m>:TBTO:ERRor:ZERO <ErrorZERO>
Enables triggering, if a zero ternary symbol pair (0-0) is detected.
Suffix:
<m>
.
1..3
Parameters:
<ErrorZERO> ON | OFF
*RST: ON
TRIGger<m>:TBTO:LENGth:CONDition <TypeOperator>
Sets the condition for the length / type. Y
ou can define an exact value or a value range.
Suffix:
<m>
.
1..3
Parameters:
<TypeOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
Protocols