Remote control commands
R&S
®
RTO6
2179User Manual 1801.6687.02 ─ 05
ZERO
SC = 0 represents a start-split (SSPLIT) transaction
DC
SC = X represents "don't care" (DC)
*RST: DC
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:STOKen <SearchName>,<TokenPacket>
SEARch:TRIGger:USB:STOKen? <SearchName>
Defines, whether a search for any token packet shall be activated or not.
Parameters:
<TokenPacket> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:TCONdition <SearchName>,<OperatorET>
SEARch:TRIGger:USB:TCONdition? <SearchName>
Sets the operator to set a specific endpoint type (ET) or an ET range. The ET values
are set with SEARch:TRIGger:
USB:TMIN and SEARch:TRIGger:USB:TMAX.
Parameters:
<OperatorET> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
The endpoint type is not relevant for the search condition.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one ET value to
be set using SEARch:TRIGger:USB:TMIN.
INRange | OORange
In range, out of range. These conditions require a range of ET
values to be set using SEARch:TRIGger:USB:TMIN and
SEARch:TRIGger:USB:TMAX.
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:TMIN <SearchName>,<ET>
SEARch:TRIGger:USB:TMIN? <SearchName>
Specifies an endpoint type (ET), or sets the start value of an ET range.