EasyManua.ls Logo

R&S RTO6 - Page 2244

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2244User Manual 1801.6687.02 ─ 05
D3CL: 3 data lanes, clock lane, low power
D4NN: 4 data lanes and no clock
*RST: D4NN
BUS<m>:DPHY:CP:PROBe <ProbeCP>
Selects the type of probe used for the clock lane.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ProbeCP> DIFFerential | SINGle
*RST: SINGle
BUS<m>:DPHY:CP:SOURce <ClockSource>
Selects the source of the clock lane.
Usually, the source is one of the analog channels. Reference and math waveforms are
available as source if the trigger source is one of the analog channels but not the serial
bus.
Alternatively
, digital channels can be used if MSO option R&S RTO6-B1 is installed.
Digital and analog channels cannot be used at the same time.
For triggering on a serial bus, analog or digital channel sources are required.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ClockSource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 |
M7 | M8 |
R1 | R2 | R3 | R4
*RST: NONE
BUS<m>:DPHY:DRATe <DataRate>
Sets a data rate.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<DataRate> Range: 1000000 to 2500000000
Increment: 1000000
*RST: 800000000
Default unit: bps
Protocols

Table of Contents

Other manuals for R&S RTO6

Related product manuals