Remote control commands
R&S
®
RTO6
2289User Manual 1801.6687.02 ─ 05
TRIGger<m>:MPHY:CREDit:MIN <CreditMin>
Specifies the Credit bit pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<CreditMin> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:CREDit:MAX <CreditMax>
Sets the end value of the Credit range if TRIGger<m>:MPHY:CREDit:CONDition is
set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<CreditMax> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:PBEGin:CONDition <CondPACPBegin>
Sets the condition for P
ACP Begin. You can define an exact value or a value range.
Suffix:
<m>
.
1..3
Parameters:
<CondPACPBegin> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:
MPHY:PBEGin:
MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MPHY:PBEGin:MIN and
TRIGger<m>:MPHY:PBEGin:MAX.
*RST: EQUal
TRIGger<m>:MPHY:PBEGin:MIN <P
ACPBEGINMin>
Specifies the P
ACP Begin bit pattern, or sets the start value of a pattern range.
Protocols