Remote control commands
R&S
®
RTO6
2374User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:PCIE:ERRC:DISP <SearchName>,<Disparity>
SEARch:TRIGger:PCIE:ERRC:
DISP? <SearchName>
Enables searching for disparity errors.
Parameters:
<Disparity>
ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:PCIE:ERRC:ECRC <SearchName>,<ECRC>
SEARch:TRIGger:PCIE:ERRC:ECRC?
<SearchName>
Enables searching for end-to-end cyclic redundancy checksum (ECRC) errors.
Parameters:
<ECRC> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:PCIE:ERRC:INVP <SearchName>,<InvalidSymbol>
SEARch:TRIGger:PCIE:ERRC:
INVP? <SearchName>
Enables searching for invalid symbol errors.
Parameters:
<InvalidSymbol>
ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:PCIE:ERRC:LCRC <SearchName>,<LCRC>
SEARch:TRIGger:PCIE:ERRC:LCRC?
<SearchName>
Enables searching for link cyclic redundancy check (LCRC) errors.
Parameters:
<LCRC> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>