EasyManua.ls Logo

R&S RTO6 - Page 2376

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2376User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:PCIE:OSET:SKIP <SearchName>,<Format>
SEARch:TRIGger:PCIE:OSET
:SKIP?
<SearchName>
Enables searching for SKP ordered sets.
Parameters:
<Format> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:PCIE:OSET:TSONe <SearchName>,<Format>
SEARch:TRIGger:PCIE:OSET:TSONe?
<SearchName>
Enables searching for training sequence 1 ordered sets.
Parameters:
<Format> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:PCIE:OSET:TSTWo <SearchName>,<Format>
SEARch:TRIGger:PCIE:OSET
:TSTW
o? <SearchName>
Enables searching for training sequence 2 ordered sets.
Parameters:
<Format> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:PCIE:TLP:ADRT <SearchName>,<AddressType>
SEARch:TRIGger:PCIE:TLP:ADRT?
<SearchName>
Sets the address type to be searched for in the transaction layer type.
Parameters:
<AddressType> ANY | X32 | X64
*RST: ANY
Parameters for setting and query:
<SearchName>