Remote control commands
R&S
®
RTO6
2487User Manual 1801.6687.02 ─ 05
Parameters:
<SourceType> CLOCk | CDR
*RST: CDR
ADVJitter<m>:SIGNal:CLKSource <SourceType>
Selects the reference source for the clock signal, if ADVJitter<m>:
SIGNal:
REFSource is set to CLOCk.
Suffix:
<m>
.
1..2
Parameters:
<SourceType> C1W1 | C1W2 | C1W3 | C2W1 | C2W2 | C2W3 | C3W1 | C3W2 |
C3W3 | C4W1 | C4W2 | C4W3 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | DIFF1 | DIFF2 | COMMON1 |
COMMON2
ADVJitter<m>:SIGNal:EDGE <EdgeType>
The clock edge sets the point in time when the state of the data signal is checked.
Suffix:
<m>
.
1..2
Parameters:
<EdgeT
ype> POSitive | NEGative | EITHer
*RST: EITHer
ADVJitter<m>:SIGNal:CLEStimation
Executes an estimation of the signal for the clock frequency
.
Suffix:
<m>
.
1..2
Usage: Event
Asynchronous command
ADVJitter<m>:SIGNal:CLFFactor <Factor>
Sets a clock multiplier factor.
Suffix:
<m>
.
1..2
Parameters:
<Factor> Range: 1 to 100
Increment: 1
*RST: 1
Jitter analysis and clock data recovery