Remote control commands
R&S
®
RTO6
2506User Manual 1801.6687.02 ─ 05
You can set separate scale values for:
●
n=1 (symbol rate)
●
n= 2 (symbol duration)
●
n= 4...16: sets only one of
fset value that is relevant for all advanced jitter com-
ponents.
●
n= 17...29: sets only one of
fset value that is relevant for all advanced noise
components.
Suffix:
<m>
.
1..2
<n> 1..29
See Table 24-20
Parameters:
<Scale> Range:
1E-15 to 1E+15
Increment: 10E-6
*RST: 0.5
Default unit: s/div
ADVJitter<m>:RESult:COMPonents<n>:OFFSet <Of
fset>
Sets the of
fset for the selected component.
You can set separate offset values for:
●
n=1 (symbol rate)
●
n= 2 (symbol duration)
●
n= 4...16: sets only one offset value that is relevant for all advanced jitter com-
ponents.
●
n= 17...29: sets only one offset value that is relevant for all advanced noise
components.
Suffix:
<m>
.
1..2
<n> 1..29
See T
able 24-20
Parameters:
<Of
fset> Range: -1E+15 to 1E+15
Increment: 1E-6
*RST: 0
Default unit: s
ADVJitter<m>:DCOMposition:SPJTable <State>
Enables the display of the periodic jitter components table.
Suffix:
<m>
.
1..2
Jitter analysis and clock data recovery