Remote control commands
R&S
®
RTO6
2513User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..2
Number of the software CDR setup
Parameters:
<InitialSync> SAMPle | DATaedge
SAMPle
The first clock edge matches the first sample of the waveform at
the left border of the display.
DA
Taedge
The first clock edge matches the first edge of the data signal.
*RST: DATaedge
Usage: Asynchronous command
CDR:SOFTware<m>:RESults <Results>
The PLL requires some time to synchronize to the phase of the data stream. Y
ou can
select when the CDR algorithm returns clock edges.
Suffix:
<m>
.
1..2
Number of the software CDR setup
Parameters:
<Results> ALL | AISYnc
ALL
All clock edges are used.
AISYnc
The clock edges of the synchronization time are discarded;
results are gathered after initial synchronization of the CDR.
Thus, meaningful TIE measurement results can be obtained.
*RST
: ALL
Usage: Asynchronous command
CDR:SOFTware<m>:PLL:ORDer <PLLOrder>
Sets the order of the PLL: first or second order. PLL of higher order can compensate
for more complex jitter behavior
.
Suffix:
<m>
.
1..2
Number of the software CDR setup
Parameters:
<PLLOrder> FIRSt | SECond
*RST: SECond
Usage: Asynchronous command
Jitter analysis and clock data recovery