Remote control commands
R&S
®
RTO6
2579User Manual 1801.6687.02 ─ 05
TRACe:REMote:MODE:FILE:STOPmode.......................................................................
2579
TRACe:REMote:MODE:FILE:
ENABle............................................................................ 2580
TRACe:REMote:MODE:FILE:FILTer...............................................................................2580
TRACe:REMote:MODE:FILE:NAME <FilePath>
Sets the directory and file name where the remote trace file is stored.
Parameters:
<FilePath> String parameter
TRACe:REMote:MODE:FILE:FORMat <Format>
Sets the file format of the remote trace file.
Parameters:
<Format> ASCii | XML
TRACe:REMote:MODE:FILE:SIZE <FileSize>
Sets the maximum size of the remote trace file.
Parameters:
<FileSize> Integer value
TRACe:REMote:MODE:FILE:STARtmode <StartMode>
Defines how the remote trace is started.
Parameters:
<StartMode> AUTO | EXPLicit
AUT
O
Starts the remote trace immediately
.
EXPLicit
Starts remote trace with TRACe:REMote:MODE:FILE:ENABle
ON
TRACe:REMote:MODE:FILE:STOPmode <StopMode>
Defines when the remote trace is stopped.
Parameters:
<StopMode> AUTO | EXPLicit | ERRor | BUFFerfull
AUTO
Ends the remote trace on device shutdown.
EXPLicit
Ends remote trace with
TRACe:REMote:MODE:FILE:ENABle
OFF
Remote trace