Measurements
R&S
®
RTO6
315User Manual 1801.6687.02 ─ 05
See also: Source and Level definition.
Hysteresis
Defines a hysteresis for the middle reference level. A rise or fall from the middle refer-
ence value that does not exceed the hysteresis is rejected as noise.
Remote command:
REFLevel<m>:RELative:
HYSTeresis on page 1561
Top outer
Defines an area above the high signal level which is still considered to be high level.
Remote command:
REFLevel<m>:ABSolute:TOTube on page 1562
MEASurement<m>:REFLevel:RESult:TOUTer? on page 1565
Top inner
Defines an area beneath the high signal level which is still considered to be high level.
Remote command:
REFLevel<m>:ABSolute:TITube on page 1562
MEASurement<m>:REFLevel:RESult:TINNer? on page 1564
Bottom inner
Defines an area above the low signal level which is still considered to be low level.
Remote command:
REFLevel<m>:ABSolute:BITube on page 1562
MEASurement<m>:REFLevel:RESult:BINNer? on page 1564
Bottom outer
Defines an area beneath the low signal level which is still considered to be low level.
Remote command:
REFLevel<m>:ABSolute:BOTube on page 1563
MEASurement<m>:REFLevel:RESult:BOUTer? on page 1564
Automatic measurements