Measurements
R&S
®
RTO6
322User Manual 1801.6687.02 ─ 05
If further settings are available for a measurement, a settings icon is shown beside the
measurement's name in the
"Active measurements"
list. Tap the icon to configure the
measurement.
Figure 8-7: Selected active measurements, some have further settings
Figure 8-8: Settings for cycle area measurement
This chapter explains all settings for amplitude/time measurements. For a description
of available measurement, see Chapter 8.2.5.1, "Overview of amplitude/time measure-
ments", on page 316.
Signal threshold.......................................................................................................... 322
Area level.................................................................................................................... 322
Pulses slope................................................................................................................323
Measured slope...........................................................................................................323
Settings for delay and phase measurements (analog sources)..................................
323
Settings for delay and phase measurements (digital sources)................................... 324
Setup/Hold measurement settings..............................................................................324
Clock ref level..............................................................................................................325
Ref level...................................................................................................................... 326
Pulse count................................................................................................................. 326
Edges slope................................................................................................................ 326
Delay to trigger measurement settings....................................................................... 326
Signal threshold
Defines a signal value that must be exceeded for the signal value to be included in the
measurement. The setting is relevant for area, time, and counting measurements.
Remote command:
MEASurement<m>:
DETThreshold
on page 1493
Area level
The reference level used to integrate the waveform. The setting is only relevant for
area measurements.
Remote command:
MEASurement<m>:
AMPTime:ALEVel
on page 1493
Automatic measurements