EasyManua.ls Logo

R&S RTO6 - Page 375

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectrum analysis
R&S
®
RTO6
375User Manual 1801.6687.02 ─ 05
3. Double-tap the editing area.
The "Formula Editor" is displayed.
4.
Delete the contents of the edit field.
5.
Tap the "More" key to display further functions in the editor.
6. Tap the required FFT function key.
7. Select the source channel.
8. Close the parenthesis.
9. Tap "Enter"
10. Select the "Scale" tab to adjust the vertical and horizontal scale of the spectrum.
11. On the "Setup" tab, select "FFT setup" to configure the spectrum. See "To config-
ure the spectrum of FFT analysis" on page 375.
To configure the spectrum of FFT analysis
By default, a suitable frequency range for the expected horizontal values according to
the resolution bandwidth is selected, in respect to performance and precision. Span
and RBW settings are coupled. If a more precise evaluation is required, for example for
postprocessing in a different application, disable the coupling and change the fre-
quency ranges and resolution bandwidth values as required.
1.
Open [FFT] > "FFT setup".
The "FFT" dialog opens.
2. In the "Setup" tab, tap the "Frequency axis" button to select the type of scaling you
want to use: linear or logarithmic.
FFT analysis

Table of Contents

Other manuals for R&S RTO6

Related product manuals