497User Manual 1801.6687.02 ─ 05
See also: Chapter 13.1.1, "Setup - general settings", on page 485
SDA, SCL
Set the waveforms of the data line (SDA) and clock line (SCL).
Usually
, the source is one of the analog channels. Reference and math waveforms are
available as source if the trigger source is one of the analog channels but not the serial
bus.
Do not combine a reference waveform with channel or math waveform because the
time correlation of these waveforms might differ.
Alternatively, digital channels can be used if MSO option R&S RTO6-B1 is installed.
Digital and analog channels cannot be used at the same time.
For triggering on a serial bus, analog or digital channel sources are required.
Remote command:
BUS<m>:I2C:SDA:SOURce on page 1676
BUS<m>:I2C:SCL:SOURce on page 1676
R/W bit
Defines if the R/W bit is considered separately or as part of the address. The setting
affects the Adress Type of the trigger conditions.
Remote command:
BUS<m>:I2C:RWBit on page 1678