Protocol analysis
R&S
®
RTO6
502User Manual 1801.6687.02 ─ 05
Adress Type
"Address"
on page 503
R/W bit
Address and data ← Trigger type
Sets the trigger to a combination of address and data condition. You can specify:
Adress Type
R/W bit
"Address" on page 503
"Position" on page 504
"Data" on page 504
The address conditions are the same as for the "Address" trigger type and "Data"
on page 504.
No Ack conditions
Selects which missing acknowledge bits is detected if the trigger type is set to "Missing
Ack".
"Address Nack"
No secondary recognizes the address.
I²C (option R&S
RT
O6-K510)