Protocol analysis
R&S
®
RTO6
555User Manual 1801.6687.02 ─ 05
Autoset
Starts software algorithms for determining the signal threshold levels and bitrate.
Remote command:
BUS<m>:FAUToset on page 1671
Check bit rate
Enables the bit rate settings for the coding standards "Manchester" and "Manchester
II". This setting is not available for "NRZ Clocked", but always enabled for the coding
standard "NRZ Unclocked", and also for triggering on signals in any coding standard.
Remote command:
BUS<m>:CMSB:BITRate:ENABle on page 1762
Bit Rate
Defines the transmission speed setting for the data signal. A bit rate definition is
optional for the coding standards "Manchester" and "Manchester II", not available for
"NRZ Clocked", but obligatory for "NRZ Unclocked" (and also for triggering on signals
in any coding standard). Default bit rate is 10 Mbps, permitted bit rates range from
300 bps to 2 Gbps.
For more details on the bit rate, see Chapter 13.6.1.1, "Special features of Manchester
coding", on page 547.
Remote command:
BUS<m>:CMSB:BITRate:VALue on page 1762
Check gap time
Enables the gap time settings (always enabled for the coding standard "NRZ
Unclocked", and also for triggering on signals in any coding standard).
Remote command:
BUS<m>:CMSB:GAPTime:ENABle on page 1762
Min gap time
Specifies a minimum gap time (idle time or timeout) between two frames. A gap time
definition is optional for the coding standards "Manchester", "Manchester II" and "NRZ
Clocked", but obligatory for "NRZ Unclocked" (and also for triggering on signals in any
coding standard). Default gap time is 10 µs, permitted gap times range from 1 ns to
1 s.
Custom: Manchester / NRZ (option R&S RT
O6-K510)