Protocol analysis
R&S
®
RTO6
778User Manual 1801.6687.02 ─ 05
See also: Chapter 13.1.1, "Setup - general settings", on page
485
Forward
Defines the forward signal settings.
Y
ou can choose between a differential and single-ended signal and set the forward
source channel for the waveform. Only analog channels are available for selection.
Remote command:
BUS<m>:HBTO:FTYP on page 2088
BUS<m>:HBTO:FDIF on page 2087
BUS<m>:HBTO:FDAP on page 2087
BUS<m>:HBTO:FDAM on page 2087
Reverse
Defines the reversed signal settings. The reverse direction is optional and can be dis-
abled.
You can choose between a differential and single-ended signal and set the reverse
source channel for the waveform. Only analog channels are available for selection.
Remote command:
BUS<m>:HBTO:RTYP on page 2089
BUS<m>:HBTO:RDIF on page 2088
BUS<m>:HBTO:RDAP on page 2088
BUS<m>:HBTO:RDAM on page 2088
Ethernet 100BASE-T1 (option R&S RT