Protocol analysis
R&S
®
RTO6
855User Manual 1801.6687.02 ─ 05
Field name / Condition Summary / Enable ← Set details
This table lists the field numbers and names in the selected frame together with a sum-
mary of the user settings of checking conditions for each field and a checkbox to
enable the checking.
It also shows the user settings of checking conditions for each field and a checkbox to
enable the checking.
Select a field in the table to specify the checking conditions for this field in the "Set
details"
 dialog.
For an overview of frames and fields, see Frame types.
Remote command: 
TRIGger<m>:USBThree:FIENable on page 2200
TRIGger<m>:USBThree:FRAMe<n>:FLD<o>:ENABle on page 2200
Data
Defines for the selected field, how a data check is executed.
"Condition"
Sets the comparison condition to a specific value or a range.
"Min"
Specifies the value or sets the start value of a range.
"Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command: 
TRIGger<m>:
USBThree:
DOPerator on page 2201
TRIGger<m>:USBThree:FRAMe<n>:FLD<o>:DOPerator on page 2201
TRIGger<m>:USBThree:DMIN on page 2201
TRIGger<m>:USBThree:FRAMe<n>:FLD<o>:DMIN on page 2201
TRIGger<m>:USBThree:DMAX on page 2202
TRIGger<m>:USBThree:FRAMe<n>:FLD<o>:DMAX on page 2202
USB 3.1 (option R&S R
T
O6-K570)