Make sure that:
In the [Bus] > "Setup" tab:
The correct "Protocol" is selected.
The data sources of the serial bus are channel signals.
In the "Trigger" > "Setup" tab:
The "Trigger on" = "Single event"
The trigger "Source" = "Serial bus".
The correct "Serial bus" is selected.
Frame Type
Selects the frame type for the USBPD analysis. For the available packet types, see
"Frame packet types" on page 862.
Y
ou can define individual checking parameters for the fields listed of the frame.
To specify these parameters, select a field from this list and open "Set details". You can
define the data and/or index operators and values, or the bit state.
The trigger instant is the last criterion that is fulfilled.
Remote command:
TRIGger<m>:USBPd:FRENable on page 2223
TRIGger<m>:USBPd:FRAMe<n>:ENABle on page 2223
USBPD (option R&S RT