EasyManua.ls Logo

R&S RTO6 - Page 932

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTO6
932User Manual 1801.6687.02 ─ 05
Type
Selects the type of condition to be searched for
. A
vailable are:
"K/D symbol"
"8 bit pattern"
"10 bit pattern"
"Expression"
"Error Condition"
Press "Define search details" to open a dialog for defining search type dependent set-
tings.
Remote command:
TRIGger<m>:EBTB:STYPe on page 2338
K/D symbol
Selects the data character (Dx.y) or control character to be searched for.
K/D Codes Symbol ← K/D symbol
Selects the data character (Dx.y) or control character to bea searched for. You can
specify the value of the data character to be searched for with "Dx Value" and "Dy
Value".
Remote command:
SEARch:TRIGger:EBTB:SYMType on page 2343
y value ← K/D symbol
Sets the y value of the data character Dx.y to be searched for, if "K/D symbol" > "Dx.y".
Remote command:
SEARch:TRIGger:EBTB:DY on page 2342
x value ← K/D symbol
Sets the x value of the data character Dx.y to be searched for, if "K/D symbol" > "Dx.y".
Remote command:
SEARch:TRIGger:EBTB:DX on page 2342
Error Condition
Searches for the selected error conditions.
8b/10b (option R&S R
T
O6-K590)

Table of Contents

Other manuals for R&S RTO6

Related product manuals