EasyManua.ls Logo

R&S RTO6 - Page 938

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTO6
938User Manual 1801.6687.02 ─ 05
Link
Selects the link width, the number of lanes that are used for the transmission of the
data.
Remote command:
BUS<m>:
PCIE:LNKW on page
2348
Descrambling
Enables descrambling of the data.
Remote command:
BUS<m>:PCIE:DSCRambling on page 2347
Lane 0/1/2/3
Select the signal sources for the logical lanes.
Usually, the source is one of the analog channels. Reference and math waveforms are
available as source if the trigger source is one of the analog channels but not the serial
bus.
Remote command:
BUS<m>:PCIE:LZER:SOURce on page 2348
BUS<m>:PCIE:LONE:SOURce on page 2348
BUS<m>:PCIE:LTWO:SOURce on page 2348
BUS<m>:PCIE:LTHRee:SOURce on page 2348
Thresholds
Sets the threshold value for digitization of signals for each line.