EasyManua.ls Logo

R&S RTO6 - Page 969

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTO6
969User Manual 1801.6687.02 ─ 05
Error type
Y
ou can search for
"Frame Incomplete" and "Length error".
Remote command:
SEARch:TRIGger:DDR:ERRor<m>:ENABle on page 2394
SEARch:TRIGger:DDR:ERENable on page 2394
Set details
In this dialog, you can define the individual search parameters for the fields of the
selected frame. Select the frame type first, and then open the "Set details" dialog.
Remote command:
SEARch:TRIGger:DDR:
FIENable on page 2394
Value ← Set details
Sets the data value of the field to be searched for. The search data setup consists of a
comparison condition and one or two data patterns.
"Condition"
Sets the comparison condition to a specific value or a range.
"Min"
Specifies the value or sets the start value of a range.
"Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
SEARch:TRIGger:DDR:
DMAX on page 2395
SEARch:TRIGger:DDR:FRAMe<m>:FLD<n>:DMAX on page 2395
SEARch:TRIGger:DDR:DMIN on page 2395
DDR (option R&S RT
O6-K91 )

Table of Contents

Other manuals for R&S RTO6

Related product manuals