In the "Settings" tab, you configure the transient response measurement parameters
and display settings.
Horizontal SMPS Frequency
Sets the SMPS switching frequency. Set the value according to your DUT
.
Remote command:
POWer:TRANsient:FREQuency on page 2555
Derived scale
Shows the scale of the displayed results. This value is calculated as the inverse of the
"SMPS switching frequency"
Edge trigger
Sets the properties of the trigger.
Source ← Edge trigger
Sets the source channel of the trigger.
Remote command:
POWer:TRANsient:TRGChannel on page 2556
Slope ← Edge trigger
Sets the edge type for the trigger event.
"Positive"
Selects the rising edge, that is a positive voltage change.
"Negative"
Selects the falling edge, that is a negative voltage change.
Output