Remote control commands
R&S
®
RTO6
1415User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
Standard Minimum value Maximum value
PAL 1 625
PAL-M 1 525
NTSC 1 263 in odd field
262 in even field
SECAM 1 625
480p/60 (P480L60HZ) 1 525
720p/30 (P720L30HZ)
720p/50 (P720L50HZ)
720p/60 (P720L60HZ)
1 750
1080i/50 (I1080L50HZ)
1080i/60 (I1080L60HZ)
1080p/24 (P1080L24HZ)
1080p/24sF (P1080L24HZSF)
1080p/25 (P1080L25HZ)
1080p/30 (P1080L30HZ)
1080p/50 (P1080L50HZ)
1080p/60 (P1080L60HZ)
1 1125
TRIGger<m>:TV:LFIeld <LineField>
The commands is only relevant for NTSC signals and sets the field in which the line
number is counted.
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
Parameters:
<LineField>
FIELD1 | FIELD2
FIELD1 = odd field
FIELD2 = even field
*RST: FIELD1
Usage: Asynchronous command
TRIGger<m>:TV:CUSTom:SCANmode <ScanMode>
Sets the scanning system. Only relevant if TRIGger<m>:TV:STANdard
is set to
CUSTom.
See also: "Scan" on page 219.